Passeo-14 PTA Balloon Catheter

Indicated for balloon dilation of the stenotic portion of a lower limb artery for the purpose of improving profusion.*

Passeo-14 product image

Product Highlights

People icon

Up to 3.8 x faster deflation times1

Coating icon

Enhanced crossability2

Arrow icon

High pushability and flexibility2


Balloon Catheter
Catheter Type OTW
Recommended guide wire 0.014”
Tip Optimized entry profile and colored
Balloon material SCP (Semi-Crystalline Polymer), controlled compliance (4 - 6%)
Balloon folding 3-fold
Balloon coating Hydrophilic patchwork coating
Balloon markers 2 swaged markers (zero profile)
Sizes ø 1.5 - 4.0 mm; L: 20 - 220 mm
Distal shaft 3.1F, hydrophilic coating, coaxial design;
150 mm length (ø 1.5 / 2.0 x 20 - 100 mm);
75 mm length (ø 2.0 x 140 - 220 mm and ø 2.5 - 4.0 mm)
Proximal shaft 3.9F, hydrophobic coating, coaxial design; stiffening wire
Usable length 150 cm (ø 1.5 - 4.0 mm); 120 cm (ø 1.5 - 2.0 mm); 90 cm (ø 2.5 - 4.0 mm)

Balloon diameter x length (mm)
ø 1.5 x 20-70 ø 2.0 x 40-220 ø 2.5 x 40-220 ø 3.0 x 40-220 ø 3.5 x 40-140 ø 4.0 x 40-140
Nominal Pressure

(NP)
atm* 7 7 7 7 7 7
ø (mm) 1.5 2.0 2.5 3.0 3.5 4.0
Rated Burst Pressure

(RBP)
atm* 14 14 14 14 14 14
ø (mm) 1.57 2.08 2.61 3.18 3.63 4.16
*1 atm= 1.013 bar

Catheter

Length (cm)
Balloon

ø (mm)
Balloon length (mm)
20 40 70 100 140 180 220
4F
Antegrade Approach
120 1.5 380271 380277 380283 - - - -
120 2 - 380278 380284 380290 380296 380302 380308
90 2.5 - 380279 380285 380291 380297 380303 380309
90 3 - 380280 380286 380292 380298 380304 380310
90 3.5 - 380281 380287 380293 380299 - -
90 4 - 380282 380288 380294 380300 - -
Catheter

Length (cm)
Balloon

ø (mm)
Balloon length (mm)
20 40 70 100 140 180 220
4F
Crossover Approach
150 1.5 380313 380319 380325 - - - -
150 2 - 380320 380326 380332 380338 380344 380350
150 2.5 - 380321 380327 380333 380339 380345 380351
150 3 - 380322 380328 380334 380340 380346 380352
150 3.5 - 380323 380329 380335 380341 - -
150 4 - 380324 380330 380336 380342 - -

IFU Link

Refer to this IFU link for a copy of the Instructions for Use and for a complete listing of the indications, contraindications, warnings and precautions.

References:

*Indication as per IFU


  1. Data on file. Volume adjustment: A 3mm x 120mm balloon contains 17% less contrast media volume than a 3mm x 140mm balloon. The measured deflation time of a 3mm x 140mm balloon was adjusted by 17 % to make a direct competitive comparison
  2. Data on file

Teleflex, the Teleflex logo, and Passeo are trademarks or registered trademarks of Teleflex Incorporated or its affiliates in the U.S and/or other countries. All other names are the trademarks or registered trademarks of their respective owners. Refer to this IFU link for a copy of the Instructions for Use and for a complete listing of the indications, contraindications, warnings and precautions. Information in this material is not a substitute for the product Instructions for Use. Not all products may be available in all countries.
© 2025 Teleflex Incorporated. All rights reserved.