Indicated for improving arterial luminal diameter in patients with clinical symptoms attributable to atherosclerotic stenosis of the renal arteries.*
Proximal gold marker for superior visibility to support accurate stent placement1
Cobalt chromium alloy combining a lower profile with high radial force1
Double helix stent design for high flexibility1
| Stent | |
|---|---|
| Stent | Balloon-expandable |
| Stent material | Cobalt Chromium (L605) |
| Strut thickness | 120 µm (ø 4.5 - 5.0 mm) |
| Strut thickness | 140 µm (ø 6.0 - 7.0 mm) |
| Stent coating | proBIO™ (Amorphous Silicon Carbide) |
| Stent marker | Proximal gold marker |
| Sizes | ø 4.5 - 7.0 mm; L: 12 - 19 mm |
| Delivery system | |
|---|---|
| Catheter type | Rapid exchange (Rx) |
| Recommended guide wire | 0.014” |
| Tip | Soft, short and tapered |
| Balloon markers | 2 swaged markers |
| Shaft (proximal) | Hydrophobic coating |
| Usable length | 140cm |
| Nominal Pressure (NP) | 10 atm |
| Rated Burst Pressure (RBP) | 15 atm (ø 4.5 - 6.0 mm) |
| Rated Burst Pressure (RBP) | 13 atm (ø 7.0 mm) |
| Balloon diameter (mm) | |||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ø 4.5 | ø 5.0 | ø 6.0 | ø 7.0 | ||||||||||||||||||||||
| Nominal Pressure (NP) |
atm* | 10 | 10 | 10 | 10 | ||||||||||||||||||||
| ø (mm) | 4.5 | 5 | 6 | 7 | |||||||||||||||||||||
| Rated Burst Pressure (RBP) |
atm* | 15 | 15 | 15 | 13 | ||||||||||||||||||||
| ø (mm) | 4.7 | 5.3 | 6.2 | 7.2 | |||||||||||||||||||||
| *1 atm= 1.013 bar | |||||||||||||||||||||||||
| Stent ø (mm) |
Catheter length 140 cm Stent length (mm) |
|||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 12 | 15 | 19 | ||||||||||||
| 4F | 4.5 | 358582 | 368711 | 358586 | ||||||||||
| 4F | 5 | 358583 | 368712 | 358587 | ||||||||||
| 5F | 6 | 358584 | 368713 | 358588 | ||||||||||
| 5F | 7 | 358585 | 368714 | 358589 | ||||||||||